Translating a VDM Model of a Medical Device into Kapture

June 11, 2025 Β· Declared Dead Β· πŸ› arXiv.org

πŸ‘» CAUSE OF DEATH: Ghosted
No code link whatsoever

"No code URL or promise found in abstract"

Evidence collected by the PWNC Scanner

Authors Joe Hare, Leo Freitas, Ken Pierce arXiv ID 2506.09636 Category cs.SE: Software Engineering Citations 0 Venue arXiv.org Last Checked 5 months ago
Abstract
As the complexity of safety-critical medical devices increases, so does the need for clear, verifiable, software requirements. This paper explores the use of Kapture, a formal modelling tool developed by D-RisQ, to translate an existing formal VDM model of a medical implant for treating focal epilepsy called CANDO. The work was undertaken without prior experience in formal methods. The paper assess Kapture's usability, the challenges of formal modelling, and the effectiveness of the translated model. The result is a model in Kapture which covers over 90% of the original VDM model, and produces matching traces of results. While several issues were encountered during design and implementation, mainly due to the initial learning curve, this paper demonstrates that complex systems can be effectively modelled in Kapture by inexperienced users and highlights some difficulties in translating VDM specifications to Kapture.
Community shame:
Not yet rated
Community Contributions

Found the code? Know the venue? Think something is wrong? Let us know!

πŸ“œ Similar Papers

In the same crypt β€” Software Engineering

Died the same way β€” πŸ‘» Ghosted